Pressure Limits: 3 x full scale differential pressure; burst: 2500 psig.
Temperature Limits: Ambient Operating: -40 to 140°F (-40 to 60°C); Process Interface: -40 to 212°F (-40 to 100°C); Storage: -40 to 212°F (-40 to 100°C).
Compensated Temperature Range: -20 to 160°F (-29 to 71°C).
Thermal Effect: ±2% F.S./50°F (reference to 77°F).
Power Requirements: 12-30 VDC for 4-20 mA outputs; 8-14 VDC for 1-5 VDC outputs, both with reverse polarity protection.
Output Signal: 4 to 20 mA DC or 1-5 VDC.
Zero and Span Adjustments: Fixed.
Response Time: 20 ms.
Loop Resistance: 900 ohms max. @ 30 VDC for current outputs. For voltage outputs, minimum lead resistance 50k ohms.
Current Consumption: 4-20 mA for current output models; 3 mA for voltage output models.