Service: Clean air and compatible, non-combustible gases. Accuracy: ±5% of full scale. Response Time (90%): 5 sec (typical). Temperature Limits: 32 to 122°F (0 to 50°C). Humidity Limit: 0-90% RH, non-condensing. Power Requirements: -A models 24 VFC +10% - 15%; -V models 24
VDC or 24 VAC +10% - 15%. Output Signal: -A models 4-20 mA current loop; -V models 0-10 VDC.
Loop Resistance: (-A models ) 700 ohms.
Current Consumption: 60 mA + output current.
Maximum Start Up Current: 85 mA; 10 V.
Output Current Limit: (-V models) >10 mA.
Electrical Connections: Screw terminal. Cable gland for 4-8 mm wire (16 gauge
wire).
Enclosure Rating: NEMA 6 (IP67) except sensing point. Probe Dimensions: 9.45 x .75" (240 x 19 mm). Mounting Orientation: Unit not position sensitive. Probe must be
aligned with air flow. Weight: 8.8 oz (250 g). Agency Approvals: CE.