Measure Temperature, Humidity, Dew Point, Pressure, or Voltage
Temperature/Humidity/Dew Point Sensor: 4" plastic temp sensor
for air only; Temp. Range: -40 to 170°F (-40 to 75°C), Accuracy: ±2°F
(±1°C); Humidity Range: 0 to 100% @ 32 to 130°F (0 to
55°C), Accuracy: ±2%; Dew Point Range: 32 to 130°F (0
to 55°C). Temperature/Pressure Sensor: 4" SS submersible temp sensor; Temp.
Range: -40 to 170°F (-40 to 75°C), Accuracy: ±2°F (±1°C);
Press. Range: 0 to 500 psi (0 to 35 bar), accuracy: ±1% FS. Temperature/Voltage Sensor: 4" SS submersible temp sensor; Temp. Range:
-40 to 170°F (-40 to 75°C), Accuracy: ±2°F (±1°C);
Voltage Range: 24 to 270 VAC, 24 to 180 VDC. Memory Size: 4,000 samples. Display: Alphanumeric backlit LCD, 16 characters x 2 lines. Sampling Method: 5 hr, 15 hr, 30 hr, 60 hr, 2 weeks, or 40 days (field
selectable). Sampling Rates: Field selectable from 5 seconds to 15 minutes. Computer Requirements: IBM compatible 286 or above and MS-DOS 3.0
or later. DB25 Male-Male. Power Requirements: 115 VAC 50/60 Hz adapter (220 VAC available) or
9V alkaline battery for 24 hr backup power source (not included). External
power may be supplied from any 12 VDC source. Printer Interface: Parallel centronics compatible printer port. Cable Length: All are 6 ft (1.8 m). Weight: 10 oz (284 g). Agency Approvals: CE.