Stainless Steel, Explosion-proof, Accuracy ±0.25%, 4-20 mA Signal
Service: Liquid, gas or vapor. Wetted Materials: 316L SS. Body: 316 SS. Accuracy: ±0.25% of calibrated span. Stability: ±0.5% of upper range limit for six months. Temperature Limits: Process Interface -40 to 212°F (-40
to 100°C). Pressure Limits: 300% full scale. Compensated Temperature: -20 to 180°F (-29 to 82°C). Thermal Effect: (includes zero and span) ±.02% upper
range/°F (30 to 130°F), ±.032% upper range/°F (-20
to 180°F). Power Requirements: 12 to 40 VDC with reverse polarity protection. Output Signal: 4 to 20 mA DC, max. 30 mA DC (2-wire). Zero and Span Adjustments: ±10% each. Loop Resistance: 600 ohms @ 24 VDC; max. ohms = (supply voltage
- 12) x 50. Electrical Connection: 1/2" female NPT. Process Connection: 1/2" female NPT. Response Time: Time constant of 20 ms. Weight: 1.67 lb (752 g). Agency Approvals: CE, FM.