Selectable Ranges of Uni-Directional or Bi-Directional, Selectable Outputs
Service: Gases or liquids compatible with 17-4 PH stainless steel. Accuracy: All pressure ranges have ±1% full scale accuracy
except the lowest selectable range of each unit is ±2% full scale. Stability: ±0.5% per year. Temperature Limits: Compensated temperature range: 32 to 130°F
(0 to 54°C); Operating temperature range: -4 to 185°F (-20 to 85°C). Pressure Limits: Max working pressure: WWDP-1: 50 psi;
WWDP-2: 100 psi; WWDP-3: 250 psi; Proof pressure: 2.2X of
full scale; Burst pressure: 40X of full scale. Thermal Effect: 2% FS/100°F (50°C) includes zero and span. Power Requirements: 12 to 30 VDC/18 to 28 VAC (Reverse
Excitation Protected). NOTE: 4-20 mA output cannot be powered
with AC voltage. Output Signal: Selectable 0-5, 0-10 and 1-5 VDC; 4 to 20 mA. Zero & Span: Digital "re" zero button (should be used when changing
ranges). Span can be adjusted by changing between field selectable
ranges. Response Time: 1 to 5 sec (selectable). Loop Resistance: 1000 ohms. Current Consumption: VDC power: 0-5, 1-5 VDC output 4 mA
(typ); 0-10 VDC output 5 mA (typ); 4-20 mA output 20 mA max.
Current consumption will equal the transmitter output in current
mode. VAC power: 0-5, 1-5, 0-10 VDC output 40 mA (typ). Electrical Connections: 1/2" conduit. Process Connections: 1/8" female NPT internal. Enclosure Rating: Designed to meet NEMA 4 (IP66). Mounting Orientation: Vertical; mount the pressure ports down
(keeps debris from building up inside the pressure port). Size: 4 x 6 x 2 in (102 x 152 x 51 mm). Weight: 1.5 lb (680.4 g). Agency Approvals: CE.